1213W RF Generator

1213W RF Generator is 1200 Watts RF Power at 13.56 MHz
for Industrial and Laboratory Applications. Water colled.

PDF DOWNLOADS:

1213 RF Generator
 
 

STANDARD FEATURES

  • 13.56 MHz up to 1200 Watts
  • Low harmonic level at 1100W h2< -50 dBc, h3 and higher < -55 dBc
  • Measuring forward, reflected and power VSWR simultaneously
  • Back Panel Control & Monitoring of all Generators functions. Data acquisition: Status Monitoring & Power Measurement via Analog Port (Basic and D-Comm models)
  • AGC Power Leveling: Output Power Control to better than +/-10W of set value.
  • Pulse operation in MGC/Burst Mode
  • Front Panel Control and Monitoring for AG 1213W models

SPECIFICATIONS

DESCRIPTION RANGE/CHARACTERISTIC
Class of Operation Class C/D
Frequency of Operation / Frequency Stability
13.56 MHz default / 0.005%
RF Power Output 1200 Watts into 50 Ohm
Output(as amplifier) in MGC/Burst Mode 0 dBm IN, 10V CTL IN pin 5 1200W ±12W
RF Input Drive for AGC Recommended +0 to +3 dBm for the best operation
Input Drive Source(amplifier) Signal or function generator, analog computer input capable of up to 2 Vp-p @ 50 Ohm
Internal RF Source Crystal oscillator at 13.56 MHz
Input and Output Impedance
50 Ohm
Input VSWR 1.2:1 max
Output VSWR 3:1 max
Output VSWR Protection 250 W max reflected power limit. Automatic foldback limit with [W] level of protection depending on actual VSWR of a load connected.
Harmonic level @ 550W
Better than - 50 dBc for 2-nd harmonic, any other > -55 dBc
Spurious Output - 55 dBm equivalent noise level generated by internal circuits
Output Blanking For pulsed applications, T&C amplifiers and generators offer blanking of the output signal for minimum noise RF spectrum
Dynamic Power Range 3 to 1200W, settings within +/- 12W
Output Settings & Control (Communications)

SubD 25 Analog on Basic model. Port Power scales: 5V and 10V (default). Rear Panel

SubD 25 Analog and Digital I/O on D-Comm models.
RS 232, 485 and USB. Rear Panel.

Front panel controls, back panel Analogue and Digital communications for AG 1213W model.

Pulse Specifications Pulse Width from 2 mikroseconds to continues, user defined
Power Margin 120% Note! End Stage Pdc margin @ 25C water temp.
RF Input / Output connectors

INPUT BNC Female
OUTPUT N Female
BLANKING BNC Female
Digital I/O (optional): RS232, USB2 and RS485

AC Power Source 200 - 240 VAC, +/-10%; 50/60 Hz, +/-10%
(100 to 120 VAC operation possible at up to 50% output level)
AC Power connection Permanently attached line cord with plug (meeting safety requirements of the region where equipment will be operating). PFC of 95% at 230 VAC.
AC Circuit Protection Back Panel Thermal Circuit Breaker, 12A
AC Input Current (RMS) RF Out 1200W: I<9A @ 220V
Cooling

Water cooled and forced air, heatsink temperature monitored for equipment safety at 70C limit.

Water flow >1 GPM (4 LPM) and temperature 20C-25 C.

Water Inlet and Outlet The connection provided accepts a 1/4-inch NPT male pipe thread.
Case

Designed to meet EMI and RFI shielding requirements aluminum natural finish.

Front Panel: T&C off white for 1213W (basic) and 1213W D-Comm models. Plastic overlay with membrane switches for AG 1213W models.

Weight 9.5 kg, 21 lbs.
Mounting 1/2" Rack, 3U high. Optional: Rack Mount Kit, Adapter Kit, Coupling Screws.
Dimensions (with hardware)

H135mm x W212mm x L482mm ( 5.25” H x8.3” x 19” )

Environmental conditions Temperature: 10° to 40° C ambient air
Humidity: 80%
Equipment intended for ISM applications in laboratory and light industrial environment.
 
About T&C | Plasma Systems | AG Amplifiers | ULTRA Amplifiers LA Amplifiers HV GenerarorsContact | Careers