STANDARD FEATURES
• 50 kHz to 800 kHz up to 500 watts CW Mode into 50 ohm load
• Input and Output Impedance 50 Ohm
• No Front Panel Controls
• Rear Panel Hardware connections for BNC Female input and
output, Blanking Control, DB25 Connector for Remote Control,
AC Input Connectors, AC Switch and Indicator, Safety Ground
Stud, Safety Labels, and Remote Control Label.
APPLICATIONS
- Ultrasonic and gas plasma application
- General laboratory use.
SPECIFICATIONS
| DESCRIPTION |
RANGE/CHARACTERISTIC |
| Class of Operation |
Class B |
| Frequency of
Operation |
20 kHz to 2000
kHz |
| RF Power Output |
Up to 500 W CW mode over frequency
range of 50 to 800 kHz, 400 W 20 to 1000 kHz, 250 W 20
kHz to 2000 kHz into 50 OHm load. |
| Gain |
57 dB @ 500W / 0.5 MHz set
point
+/- 1.5 dB 30 kHz to 800 kHz |
| RF Power Input Drive for
AGC |
Recommended - 2 dBm to 0 dBm |
| RF Input Drive |
-15 dBm to 0 dBm typical range,
Up to +3 dBm max |
| Input Drive Source |
Signal or function generator,
analog
computer input capable of up to 2
Vp-p @ 50 Ohm |
| Input and Output
Impedance |
50 Ohm (Optimized
for 50 Ohm !) |
| Input VSWR |
2:1 max |
| Output VSWR |
3:1 max |
| Output VSWR Protection |
100 W max reflected power limit |
| Harmonic Level @ 300W |
-17 dBc or better for 3-d
- 22 dBc or better for any other |
| Harmonic Level @ 100W |
- 17 dBc or better for 3-d
- 22 dBc or better for any other |
| Spurious Output |
- 50 dBc equivalent noise level
generated by internal circuits |
| Output Blanking |
For pulsed applications,
ULTRA 06-800
amplifier offers blanking of the output
signal for minimum noise RF
spectrum.
“T off” time < or = 10 microsecond,
“T on” time < or = 10 microsecond,
TTL Low - RF Output Blanked,
TTL High - RF Output present, normal
operation.
BNC Female connector. Back Panel |
| RF Connectors |
BNC Female: Back Panel |
| Third Order Intercept |
+57 dBm |
| AC Power Source |
100 - 120, 200 - 240 VAC, +/- 10 %, 50/60 Hz
100 - 120V, 11A, 200 - 240 V, 5.5A
|
| Panels |
Front: blank panel, air intake.
Marking:
company name, product number.
Back: air exhaust. Hardware location
for RF IN, RF OUT, Blanking Control,
DB25 Connector for Remote Control, AC Input connector,
AC Switch and Indicator,
Fan Guard, Safety Ground Stud.
Markings: safety labels, GND and
AC Line Table, Serial Number and
Remote Control Label. |
|
|